Veterinary Mobile Anesthesia Machine YSAV602V
Anesthesia Machine, designed with state-of-the-art technologies, is one of high utility with excellent performance and complete functions. Both outstanding and inexpensive, it has the best cost performance that can satisfy any customer.
1. The Ventilator adopts microcomputer control with high-definition 5.7” LCD display,integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
2. The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
3, Reliable pneumatic transmission system.
4. Economic and proper configuration of technologies, optional as required by the Customer.
5. Modularized design and upgrading. Upgrading is available as required by the Customer.
Specification
Main body
High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application
adult, child
Gas source
O2: 0.27 ~ 0.55MPa N2O: 0.27 ~ 0.55MPa
Flow meter
O2: 0.1~10L/min N2O: 0.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply
25~75L/min
Low oxygen pressure alarm
There will be sound alarm when the oxygen
pressure < 0.2MPa
Vaporizer
It has the function of automatic compensation
based on pressure, temperature, and flow rate.
The regulation range of evaporator
concentration is 0~5 vol%. Among
Halothane, Enflurane, Isoflurane and
Sevoflurane, one can be chosen for application
as required by the customer. And vaporizers
that are imported with original packaging are
also available.
Respiratory circuit
working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
Respiratory bellows
bellows for adults, bellows for children
tidal volume range: 0 ~ 1,500 ml
Ventilator
Display mode
High-definition 5.7” LCD screen display
Ventilation mode
IPPV, SIPPV, SIMV, PEE,MANUAL, SIGH
Ventilation parameters
Tidal volume
50~1500ml
Rate
2~99bmp
SIMV rate
2~20bpm
I:E
2:1~1:8
Inspiratory trigger pressure
-1.0~2.0 kPa
PEEP
0~2.0 kPa
Pressure range
1.0~6.0 kPa
SIGH
1.5 times the inspiratory time
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure
Monitoring of oxygen concentration
(21%~100%)
Security Alarm System
Oxygen Concentration alarm
upper limit setting range (50%~100%)
low limit setting range (15%~50%)
Airway pressure alarm
upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm
it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm
it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
Power
AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
Items within brackets are optional.

Specifications are sourced from the authorized supplier catalog and must be confirmed for the quoted configuration. Regulatory documentation is verified by model and destination country.